• 浩瀚 三氟一溴甲烷灭火剂1301测定GDX-104填充柱3m*4mm

    详细信息

     品牌:浩瀚  型号:3m*4mm  加工定制:是  
     测量范围:ppm  测量对象:三氟一溴甲烷  控温范围:300 ℃ 
     尺寸:1500 mm 重量:0.02 Kg 碳分子筛:TDX-01  
     TDX-02:苯系物  医疗器械:环氧乙烷  甲烷:甲醇  
     赛默飞:安捷伦  甲酸:消毒剂  :氙  
     水质 五氯酚:玻璃柱  三氟一溴甲烷:1301灭火剂   

    三氟一溴甲烷灭火剂(1301灭火剂) GB6051-1985

    3.2.1.2   测定条件 
             a.  色谱柱:3m长不锈钢管,内径为4mm; 
             b.  固定相:GDX-104,60~80目。使用前需活化,活化条件为通氮气(30~50ml/min ), 140℃不少于6h; 
             c.  载气:氢气,流量为45ml/min (用皂膜流量计实测); 
             d.  温度控制:柱温为115℃,汽化温度为100℃,检测室温度为100℃; 
             e.  桥电流:200mΑ; 
              f.  记录仪:满标量程为5mV ,满标长度为250mm,纸速为10mm/min; 
             g.  检测器灵敏度的控制:为确保色谱分析灵敏度,在规定的操作条件下,以1301为样品,灵敏度S1301应不低于1500mV·ml/ml,由式(1)表示。 

    式中:1.065――峰面积校正系数; 
                     A――1301色谱峰面积,m㎡; 
                   C1――记录纸速的倒数,min/mm; 
                   C2――记录仪灵敏度,mV/mm ; 
                     F――载气流量,ml/min ;  
                     V――1301进样量,ml; 
                     K――检测器衰减倍数。 

     
     
     
     
     
     
     
     

    HJ 591-2010 Water quality Determination of pentachlorophenol Gas chromatography

    HJ 591-2010 Water Quality Determination of Pentachlorophenol Gas Chromatography was implemented on January 1, 2011.

    This standard replaces GB/T 8972-1988 Water Quality Determination of Pentachlorophenol by Gas Chromatography.




    Name: Agilent 7890B glass column



    Specification: 2m * 1/4, 80-100 meshes



    Model: 1.5% OV-17+2% QF-1/Chromosorb AW DMCS (80-100 meshes)



    No.: 20220818001



    Maximum temperature: 240 ° C



    Application: HJ 591-2010 Water Quality Determination of Pentachlorophenol Gas Chromatography



    3.2.1.2 Determination conditions

             a. Chromatographic column: 3m long stainless steel tube, with inner diameter of 4mm;  

             b. Stationary phase: GDX - 104, 60 ~ 80 meshes. Activation is required before use, and the activation condition is nitrogen (30~50ml/min), 140 ℃, no less than 6h;  

             c. Carrier gas: hydrogen, flow rate 45ml/min (measured with soap film flowmeter);  

             d. Temperature control: the column temperature is 115 ℃, the vaporization temperature is 100 ℃, and the detection room temperature is 100 ℃;  

             e. Bridge current: 200m Α;  

    f. Recorder: full scale range is 5mV, full scale length is 250mm, and paper speed is 10mm/min;  

             g. Control of detector sensitivity: to ensure the sensitivity of chromatographic analysis, the sensitivity S1301 should not be less than 1500mV · ml/ml with 1301 as the sample under the specified operating conditions, expressed by formula (1).  



    Where: 1.065 - peak area correction coefficient;  

    A ―― 1301 chromatographic peak area, m ㎡;  

    C1 -- reciprocal of recording paper speed, min/mm;  

    C2 ― recorder sensitivity, mV/mm;  

    F ― carrier gas flow, ml/min;  

    V-1301 injection volume, ml;  

    K ---- attenuation multiple of detector.