GDX-502 detected in the application of non-methane hydrocarbons
Size: 2m * 3mm
Material: Stainless steel
Packing; GDX-502
Mesh: 60-80
Column temperature: 60 ° C
The gasification chamber: 140 ° C
Detector: 180 ° C
Carrier gas: N2
Flow rate: 30ml / min 0.01MPa
Injection: 1ml
Sensitivity: 1
Reference spectrum:
Instructions for use:
Aging factory before, after a period of use for some time, such as baseline instability, re-aging.
The column inlet end (long end) connected to the inlet, and the other end is not connected to the detector.
High purity nitrogen gas through a flow rate of 40ml / min, 50 ° C initial temperature to 5 ° C / min temperature programming to 220 ° C, constant temperature for 12 hours.
After cooling, you can use the other end to the detector.
Aging Well then, should achieve stable baseline separation is good.